ไฮไลท์
VICTOR 07, Temperature Calibrator Multi-function process calibrator thermocouple calibrator,
Name : Temperature Calibrator Multi-function,
Brand : VICTOR ,
Model: VICTOR 07,
ข้อมูล
น้ำหนัก
บาร์โค้ด
ลงสินค้า
อัพเดทล่าสุด
รายละเอียดสินค้า

VICTOR 07 Temperature Calibrator Multi-function process calibrator thermocouple calibrator

Products Description

VICTOR 07 Temperature Calibrator with highest accuracy is 0.05% displayed as 5 digits

 

1.Input measurement

Function: DC voltage, DC current, resistance, thermocouple,thermistor, on/off

Measurement accuracy of 0.05%, measurement display of5 digits

Measurement rate: 2-3 times/second

OHM and RTD can perform 2W/3W/4W measurements

The measured temperature value and corresponding temperature signal can be switched for display

Loop detection function: internal 24V circuit power supply and simultaneous measurement of current

mA has a percentage display mode

2.Analog output

Function: DC voltage, DC current, resistance, thermocouple,thermistor

Output accuracy of 0.05%, output display of 5 digits

Can manually increase or decrease output settings by position

OHM excitation current has two settings: H/L

The output temperature value and corresponding temperature signal can be switched for display

DCl output includes active output and passive output

mA has a percentage display mode

When outputting DCl, manual stepping and automatic waveform output functions can be provided

3.internally provide 24V circuit power supply

4. Built in high-precision temperature sensor, capable of high-precision cold end compensation

5.DCl temperature linear conversion function

6.Allow simultaneous measurement and output of process signals, as well as simultaneous display of input measurement values and output settings

 

 

Specification

 

Input measurement function

Function

Range

Measuring range

Resolution

Accuracy

DCV

100mV

-110.000mV~110.000mV

0.001mV

0.05%+0.03mV

1V

-1.1000V~1.1000V

0.0001V

0.05%+0.3mV

35V

-35.000V~35.000V

0.001V

0.02%+2mV

DCI

30mA

-35.000mA~35.000mA

0.001mA

0.02%+4uA

ОHM

500Ω

0~550.00Ω

0.01Ω

0.05%+0.2Ω

5kΩ

0~5.5000KΩ

0.0001KΩ

0.05%+2Ω

TC

R

0℃~1767℃

1℃

2℃

S

0℃~1767℃

1℃

2℃

K

-100.0℃~1372.0℃

0.1℃

0.8℃

E

-50.0℃~1000.0°C

0.1℃

0.9℃

J

-60.0℃~1200.0°C

0.1℃

0.7℃

T

-100.0℃~400.0℃

0.1℃

0.7℃

N

-200.0℃~1300.0℃

0.1℃

0.9℃

B

600℃~1820℃

1℃

1.4℃

A

0℃~2500℃

1℃

2℃

C

0℃~2310℃

1℃

2℃

D

0℃~2310℃

1℃

2℃

RTD(3W)

Pt100(385)

-200.0℃~800.0℃

0.1℃

0.5℃

Pt200(385)

-200.0℃~630.0℃

0.1℃

0.8℃

Pt500(385)

-200.0℃~630.0℃

0.1℃

0.4℃

Pt1000(385)

-200.0℃~630.0℃

0.1℃

0.3℃

Cu50

-50.0℃~150.0℃

0.1℃

0.7℃

Continuity detection

500Ω

≤50Ω will sound

0.01Ω

 

LOOP

24V

   

10%

 

Analog output function

Function

Range

Measuring range

Resolution

Accuracy

DCV

100mV

-10.00mV~110.00mV

0.01mV

0.05%+0.03mV

1V

-0.1000V~1.1000V

0.0001V

0.05%+0.3mV

10V

-1.000V~11.000V

0.001V

0.05%+2mV

DCI

20mA

0.000mA~22.000mA

0.001mA

0.05%+4uA

ОHM

400Ω

0~400.0Ω

0.1Ω

0.05%+0.2Ω

4KΩ

0~4.000Ω

0.05%+2Ω

TC

R

0℃~1767℃

1℃

1.2℃

S

0℃~1767℃

1℃

1.2℃

K

-200.0℃~1372.0℃

0.1℃

0.5℃

E

-200.0℃~1000.0℃

0.1℃

0.4℃

J

-200.0℃~1200.0℃

0.1℃

0.5℃

T

-250.0℃~400.0℃

0.1℃

0.6℃

N

-200.0℃~1300.0℃

0.1℃

0.7℃

B

600℃~1820℃

1℃

1.1℃

A

0℃~2500℃

1℃

2.0℃

C

0℃~2310℃

1℃

2.0℃

D

0℃~2310℃

1℃

2.2℃

RTD(3W)

Pt100(385)

-200.0℃~800.0℃

0.1℃

0.3℃

Pt200(385)

-200.0℃~630.0℃

0.1℃

0.8℃

Pt500(385)

-200.0℃~630.0℃

0.1℃

0.4℃

Pt1000(385)

-200.0℃~630.0℃

0.1℃

0.2℃

Cu50

-50.0℃~150.0℃

0.1℃

0.6℃

 

VICTOR 07, Temperature Calibrator Multi-function process calibrator thermocouple calibrator,
VICTOR 07, Temperature Calibrator Multi-function process calibrator thermocouple calibrator,
VICTOR 07, Temperature Calibrator Multi-function process calibrator thermocouple calibrator,
VICTOR 07, Temperature Calibrator Multi-function process calibrator thermocouple calibrator,
VICTOR 07, Temperature Calibrator Multi-function process calibrator thermocouple calibrator,
VICTOR 07, Temperature Calibrator Multi-function process calibrator thermocouple calibrator,
เงื่อนไขอื่นๆ
Tags

วิธีการชำระเงิน

บมจ. ธนาคารกสิกรไทย สาขาแฟชั่นไอส์แลนด์ รามอินทรา ออมทรัพย์
ฺBrand Product ------------------
              
CHINCAN  
รายการสั่งซื้อของฉัน
เข้าสู่ระบบด้วย
เข้าสู่ระบบ
สมัครสมาชิก

ยังไม่มีบัญชีเทพ สร้างบัญชีใหม่ ไม่มีค่าใช้จ่าย
สมัครสมาชิก (ฟรี)
รายการสั่งซื้อของฉัน
ข้อมูลร้านค้านี้
ร้านBNCTOOLINGSHOP
BNCTOOLINGSHOP
นำเข้าและจำหน่าย,เครื่องจักร,เครื่องมือช่าง,ซ่อมบำรุง,เครื่องมือวัด,ทางอุตสหกรรม,วัดอุณหภูมิ,วัดความชื้น,วัดความดัน,วัดด้านมิติ,วัดด้านเคมี,ระดับน้ำดิจิตอล,digital level,วัดความหนา,วัดเฉดสี,วัดแรงดึงแรงกด,วัดความแข็ง,และด้านมวล,การสอบเทียบ
เบอร์โทร : 029188020
อีเมล : bnctooling@hotmail.com
ส่งข้อความติดต่อร้าน
เกี่ยวกับร้านค้านี้
สินค้าที่ดูล่าสุด
ดูสินค้าทั้งหมดในร้าน
สินค้าที่ดูล่าสุด
บันทึกเป็นร้านโปรด
Join เป็นสมาชิกร้าน
แชร์หน้านี้
แชร์หน้านี้

TOP เลื่อนขึ้นบนสุด
พูดคุย-สอบถาม