Specifications
● Calibration: According SCHMIDT factory procedure
● Accuracy: For PA from 5 % up to 100 % FS*:± 0.5 % FS* ± 1 digit remaining tension range and other calibration materials: ±3 % full scale and ±1 digit
● Memory for material curves: 2 for SCHMIDT calibration plus 9 for customized calibrations
● Measuring units: Force (switchable): cN, daN, g, kg, N, lb
● Thickness (switchable): mm, inch
● Overrange: Approx. 10 % FS*, full scale, without accuracy guarantee
● Overload protection: 100 % full scale
● Measuring principle: Strain gauge bridge
● Signal processing: Digital, 16 bit A/D converter
● Measuring frequency: Max. 1 kHz (1000 measurings/sec.), intern 8 kHz
● Measuring roller deflection: Max. 0.2 mm
● Display: Graph LCD
● 3 different displays: Numeric,
● Numeric with graph (time-tension)
● X-Y-diagram
● Display update rate: Approx. 2 times per second
● Memory: MIN, MAX, PEAK, AVG and LAST
● Output signal: USB
● Memory: 60.000 values split up to 255 measuring positions
● Memory modes: 5 memory modes with statistical evaluation
● Damping: Adjustable electronic damping (averaging)
● Cal. Adjustment: ± 10 steps in 1 % increment
● Thickness compensation: Max. 2.5 mm (not available for all models)
● Auto power off: Approx. 3 minutes of non-use
● Temperature coefficient: Gain: less than ± 0.01 % FS* /°C
● Zero point: Less than ± 0.03 % FS* /°C
● Temperature range: 10 – 45 °C
● Air humidity: 85 % RH, max.
● Power supply: LiPo accumulator (approx. 40 h continuous use), charging time approx. 3 1/2 h,
● USB AC adapter 100 – 240 V AC with 4 adapters (EU/USA/UK/AUS-NZ)
● Housing material: Die-cast aluminium
● Housing dimensions: See Dimensions
● Weight, net (gross): Up to DTX-50K approx. 875 g / 1550 g
● DTX-60K-V1 approx. 1040 g / 2700 g